38 Replacing Stand Boards
MAMMOMAT Novation DR SPB7-250.841.01.04.02 Siemens AG
05.05 CS SD 24
Page 38 of 64
Medical Solutions
D750 master board for Mammomat Novation 0
Fig. 13: D750
• Turn the system OFF.
• Remove the right stand cover.
• Remove board D750 from the stand.
• IC 45, IC 53, IC 62, IC 52 have to be moved from the old to the new
board(1/Fig. 13 / p. 38).
• Check the position of switch S5 and the bridge on …X754 1-2 (2/Fig. 13 / p. 38).
• Install the new board.
• Install the right stand cover.
• Turn the system ON.
• No backup/restore or adjustment needed.