48 Collimator and D814
MAMMOMAT Novation DR SPB7-250.841.01.04.02 Siemens AG
05.05 CS SD 24
Page 48 of 64
Medical Solutions
• Pull out the lamp holder carefully. Be careful with the lamp power supply cable.
Fig. 26: Lamp holder
• Replace the lamp (1/Fig. 26 / p. 48)
NOTE
Do not touch the lamp with your fingers. Wear gloves or use a soft
cloth when replacing the lamp.
• Reinstall the lamp holder
NOTE
Be careful not to damage the filter disk.
• Reinstall the plate with the screws (1/Fig. 24 / p. 47)
• Make an exposure to check the X-ray/light field
• If necessary, adjust the light field according to System Start-up with WH AWS,
SPB7-250.815… the Collimator Adjustment and Collimator Field Light Calibration
chapters.
Final steps 0
• For information on backing up the stand data, see Software System Stand,
SPB7-250.816.03…, Adjustment and Service Programs.
• Install the covers of the swivel arm.