50 Grid
MAMMOMAT Novation DR SPB7-250.841.01.04.02 Siemens AG
05.05 CS SD 24
Page 50 of 64
Medical Solutions
8-
7Grid
Mechanical works and adjusting 0
General 0
These instructions describe how to disassemble the grid and mount a new one. It then
describes how to perform subsequent required tasks, mainly checks of several adjust-
ments.
Documents 0
The type of documentation depends on the acquisition workstation in use:
• DROC
- Installation Instructions and Start-up (DROC)
- QC Manual (DROC)
• WH AWS
- Start-up, WH AWS VAXXX(XXX - Depends on the installed software version)
- Software, WH AWS VAXXX (xxx - Depends on the installed software version)
- QC Manual (WH AS)
NOTE
Note: removing the grid requires certain mech. adjustments. After
assembling the grid, image quality should be checked according
to the instructions Mammomat Novation, System
Start-up...,SPB7-250.815…for systems with WH AWS or Mammo-
mat Novation, Installation Instructions… SPB7-250.812…for sys-
tems with DROC.
Disassembling the grid 0
The disassembly of the grid is described in the instructions " Replacement of parts"
SPB7-250.841…, Repl.. of the FFDM(Disassembling the old detector / p. 12).