Siemens AG SPB7-250.841.01.04.02 MAMMOMAT Novation DR
05.05 CS SD 24
Collimator and D814 45
Page 45 of 64
Medical Solutions
Fig. 22: D814
Assembling board D814 0
• Plug D814 into the 4 plastic holders (1/Fig. 22 / p. 45).
• Plug connectors X833, X834, X 835, X836 and X838 on D814 (Fig. 21 / p. 44).
• Insert the shielding plate on board D814 (1/Fig. 20 / p. 44).
Mounting the collimator 0
• Lead the filter motor cable through the hole in the plate.
• Mount the collimator with board D814 on the tube assembly support using the 4 screws
M5x14. ((1/Fig. 18 / p. 43) and (1/Fig. 19 / p. 43)).
• Plug connectors X894/X839 into board D805 and X832/X837 into D814.
• Secure the cables via a cable tie. (3/Fig. 19 / p. 43)
Filter offset value 0
The filter offset value has to be checked after a collimator change. The correct value
ensures the correct position of the filter on the collimator.
For this setting perform the following:
• Connect the service PC according to Software System Stand SPB7-250.816.03…
• Start the service program
• Select Configuration->Filter->Filter offset value
• The filter offset value has to be between 70 and 90