Collimator and D814 43
Page 43 of 64
Medical Solutions
Fig. 18: Tube assembly with collimator
• Turn the system OFF.
• Remove the covers from the swivel arm.
Fig. 19: Tube assembly support
• Cut the cable tie (3/Fig. 19 / p. 43).
• Remove the cables between the collimator and D805. For this, unplug connectors
X894/X895 on board D805 and connector X832/X837 on board D814 (Fig. 19 / p. 43).
• Remove the 4 screws M5x14 ((1/Fig. 18 / p. 43) and (1/Fig. 19 / p. 43)). The collimator
can be removed from the tube assembly support (Fig. 20 / p. 44).