Siemens AG SPB7-250.841.01.04.02 MAMMOMAT Novation DR
05.05 CS SD 24
Compression Unit 29
Page 29 of 64
Medical Solutions
5-
4Compression Unit
Replacing the compression unit 0
General 0
These instructions describe the work steps for replacing the compression unit and for cal-
ibrating this unit.
The procedure should be used also in cases when the compression unit has been cali-
brated.
Required documents 0
Software System Stand, SPB7-250.816.03....
QC manual
Tools 0
Standard tool kit
Dynamometer or scale with a digital display
Service PC
Disassembling the compression unit 0
• Turn the system OFF.
• Remove the covers from the swivel arm.
• Disconnect connectors X 872, X 873,X874 and X831 from board D805.
• Remove the bolts holding the compression unit.
• The compression unit can be dismounted carefully from the holder.
Compression unit assembly 0
• Install the compression unit in the reverse order of disassembly.
• Plug connectors X872,X873,X874 and X831 into board D805.
• Insert the three bolts and tighten them slightly (1/Fig.9/p.30).
• Secure the cables with cable ties.