General 0
These instructions describe the work steps for replacing the compression unit and for cal-
ibrating this unit.
The procedure should be used also in cases when the compression unit has been cali-
brated.
Required documents 0
Software System Stand, SPB7-250.816.03....
QC manual
Tools 0
Standard tool kit
Dynamometer or scale with a digital display
Service PC
Disassembling the compression unit 0
Turn the system OFF.
Remove the covers from the swivel arm.
Disconnect connectors X 872, X 873,X874 and X831 from board D805.
Remove the bolts holding the compression unit.
The compression unit can be dismounted carefully from the holder.
Compression unit assembly 0
Install the compression unit in the reverse order of disassembly.
Plug connectors X872,X873,X874 and X831 into board D805.
Insert the three bolts and tighten them slightly (1/Fig.9/p.30).
Secure the cables with cable ties.