30 Compression Unit
MAMMOMAT Novation DR SPB7-250.841.01.04.02 Siemens AG
05.05 CS SD 24
Page 30 of 64
Medical Solutions
Fig. 9: Fixing bolts
NOTE
Ensure that the compression unit is installed in the correct me-
chanical position. Use phantom images to check the unit's posi-
tion after adjustment.
NOTE
When positioning the exchange compression unit, make sure that
the two shrunk-on insulating hoses attached to the backside of the
electromotor do not become entangled in the upper part of the arc
(or dome, also called “church”). The electromotor has to move
freely under that arc. Several manual checks are necessary
(1/Fig. 10 / p. 31).