Siemens AG SPB7-250.841.01.04.02 MAMMOMAT Novation DR
05.05 CS SD 24
Compression Unit 31
Page 31 of 64
Medical Solutions
Fig. 10: Church
Insert the compression plate into the compression unit. Manually move the compres-
sion plate downward toward the detector. It is important for the front edge of the com-
pression plate to be parallel to the carbon-fiber plate edge.
Manually move the compression plate upward. There must be 285 mm between the
carbon-fiber plate edge and the metal sheet under the guide for the compression plate
holder (Fig. 11 / p. 32).
Securely tighten the three bolts for the compression unit holder (1/Fig.9/p.30).
NOTE
This document describes an initial calibration. Therefore, it de-
scribes cases in which a Mammomat system is equipped with a
newly mounted compression unit, and therefore either the relevant
data from another (e.g. the formerly mounted) compression unit
stored in the system or no meaningful compression calibration
data stored in the system.