18 Replacing basic unit components
MAMMOMAT Novation DR SPB7-250.841.01.04.02 Siemens AG
05.05 CS SD 24
Page 18 of 64
Medical Solutions
7. Locate the detector-specific files in the directory C:\aws\drul\data - they are:
MPwxyz.cal
MPwxyz.cfg
MPwxyz.map
MPwxyz.smj
qualify.log
8. Select (highlight) and cut these files from the directory C:\aws\drul\data and paste
them into the newly created subdirectory MPwxyz. (or copy the 4 files from the directo-
ry C:\aws\drul\data, paste them into subdirectory MPwxyz, and then delete them from
C:\aws\drul\data.)
9. Insert the supplied CD-ROM into the compact disk drive of the workstation.
10. Open the directory on the CD-ROM and locate the three detector-specific files for the
new detector:
mpabcd.cfg
mpabcd_final.map
mpabcd.smj
There should be three files. The .cal file is not present. It is first generated on the
MAMMOMAT Novation
DR
system during one of the following work steps, i.e., the
detector gain calibration.
11. Copy these three detector-specific files from the CD-ROM into C:\aws\drul\data using
drag and drop.
12. Rename MPabcd_final.map to MPabcd.map.
In File Manager, simply click the file name and change it from MPabcd_final.map to
MPabcd.map.
13. The MP in the file names must be in upper case letters. The file names on older CDs
were in lower case letters only. You must rename such files to e.g. MPabcd.cfg, MPab-
cd.map and MPabcd.smj. In File Manager, click the file name and replace the lower
case mp with the upper case MP.
Repeat the procedure for each file, so that all contain an upper case MP. Select
(highlight) and rename all three files.
14. Leave Windows Explorer with exit.
15. Restart the syngo application by going to Utilities in the service software and select-
ing Source > System.
Utilities > System > Restart Application
Select Restart Application and press Go to restart the syngo application.
16. When finished, verify that the code dr appears on the MAMMOMAT Novation
DR
con-
sole, which indicates that "digital radiography" mode is ready.
The detector-specific files have been installed.
Checks, adjustments and calibrations
Since the detector has been replaced, some checks must be performed because the
mechanical correlation of the tube, collimator and the detector may have changed.
The procedures vary depending on the workstation installed.