Siemens AG SPB7-250.841.01.04.02 MAMMOMAT Novation DR
05.05 CS SD 24
Replacing basic unit electrical assemblies 27
Page 27 of 64
Medical Solutions
NOTE
The "time zone" link has been added in revision 05 (software ver-
sion 1.2.6.0).
If you have an earlier revision, you cannot set the time zone. The
BRICK will work fine regardless.
2. Click Modify Settings to change the time zone setting.
3. Enter the new string in the New Time Zone Value dialog box and click Change time
zones to save the new setting.
4. Go back to the BRICK home page.
Setting the xrayconfig timeout
1. Establish a connection to the BRICK via the command line interface; see "Establish-
ing a connection to the BRICK command line interface" on page 3 - 2
2. Enter the command:
brick# xrayconfig host_timeouts save 1000
Setting the date, time
1. To check the current date and time setting on the BRICK, use the following command:
brick# date
To set the date and time on the BRICK, use the following command:
brick# date mmddHHMMyy
mm - month
dd - day
HH - hours
MM - minutes
yy - year
The following example sets the system to 3:10 p.m. on November 21, 2004:
brick# date 1121031004
Rebooting
1. Reboot the BRICK by entering the command:
brick# reboot
Wait approximately 5 minutes for the reboot to complete. Check that all BRICK LEDs
are working; see "Start up the system" on page 3 - 5.