Installation 5 - 13
Siemens AG SPB7-230.814.01 Page 13 of 24 Opdima
Medical Solutions Rev. 01 10.04 CS PS 24 Installation
6. Connect the MO unit SCSI cable to the workstation SCSI connector (2C/Fig. 17).
7. Connect the MO unit cable to one of the two connectors of the MO unit. Fit the
terminator to the other connector of the MO unit.
For switch settings, SCSI ID and termination see Paragraph “Installing the MO
unit” in Chapter “Removal and replacement of sub-assemblies” in
Mammomat 1000/3000/3000 Nova - Opdimar Service Instructions.
8. Connect the mouse to the keyboard.
9. Connect keyboard and mouse (3/Fig. 17).
10. Connect monitor cable to the workstation connector (6/Fig. 17).
11. Connect Ethernet cable (4/Fig. 17).
12. Mark the wall socket connected to the same branch circuit as the MAMMOMAT
(Fig. 14) with label according to Fig. 15.
The power distribution to the system is shown in Fig. 14.
Make sure that the SCSI cables are properly connected and fixed
to the units and that the cable is not bent too much just behind
the MO unit. Also, check that the MO unit is terminated correctly.
If disregarded, this might generate a lot of problems e.g. failure to
write images to MO disk, corrupted MO disk or corrupted hard
disk.
CAUTION
Work-
station
1-phase
L1L2
N
On
Off
PE
L1 N
PE
On
Off
Power On-Off breaker
Power connection
in MAMMOMAT 3000-stand
10A
10A
Power connection
in MAMMOMAT 3000-stand
Power On-Off breaker
2-phase
Work-
station
OPD00166
Fig. 14 Power distribution