Opdima SPB7-230.814.01 Page 14 of 24 Siemens AG
Installation Rev. 01 10.04 CS PS 24 Medical Solutions
5 - 14 Installation
13. If a voltage selector switch exists, ensure the line voltage selector switch is set to
the appropriate setting: 115V or 230V.
14. Connect the AC power cords of the main unit, the monitor, the MO unit, the local
Opdimar printer (if present) and the MAMMOMAT printer (if present) to the wall
socket marked with label.
It is advisable to use an extension cord with at least three outlets.
Plugging a 115V power cord into a 230V connector will severely
damage the system.
The workstation and the printer must be grounded to the same
potential as the MAMMOMAT.
Connect the wall socket for the workstation and printer power
supply to the same branch circuit as the MAMMOMAT power sup-
ply.
(The workstation table can be connected to any branch circuit.)
CAUTION
CAUTION
OPD00600
Opdima mains power
supply outlet!
Fig. 15 Mains marking label