Siemens AG SPR8-230.831.30.03.02 MOBILETT XP Digital
11.05 CS SD 24
Inspection and maintenance 45
Page 45 of 74
Medical Solutions
3.2.7.4 Adjusting screw and cantilever
WARNING
Uncontrolled system movements pose danger of injury.
The support can fall if the adjusting screws are unscrewed incor-
rectly!
¹ Carelessness can result in death or serious bodily injury.
SIM Lock nuts
• Verify that the lock nuts on the adjusting screw are properly seated (Fig. 11 / p. 45).
SIM Attaching the base
• Verify that the base attachment in the chassis is secure (Fig. 11 / p. 45).
Fig. 11: Adjusting screw and cantilever attachment
3.2.7.5 Cable harness for the arm system
SIM Undamaged
• Check all support arm cabling for damage (stretching, twisting, insulation cracks). Pay
special attention to the stress points of moving elements (Fig. 12 / p. 46).
• Replace any damaged cables.