Repair, maintenance and servicing
6.6 Removing faults
OXYMAT 64
112 Operating Instructions, 06/2008, A5E00880383-03
6.6.4 Other faults
In addition to the fault messages which can be shown in the logbook, the following influences
could lead to an unsteady display:
Table 6- 4 Additional fault causes
Cause Remedy
Pressure surges or variations in the sample gas outlet Route the sample gas outlet separate from the outlets from
other analyzers, and/or install a damping unit in the sample
gas outlet. The flow resistance must be kept as small as
possible. Dimension the flow resistance such that the
sample gas pressure does not exceed 100 hPa (rel.).
Occurrence of sporadic spikes See also function 76; inform service department if
necessary
Green LED on the rear of the analyzer flashes in a certain
sequence (not regular flashing)
Inform Service