Operation
5.4 Operating functions
OXYMAT 64
Operating Instructions, 06/2008, A5E00880383-03
73
5.4.5.4 On/off functions (function 52)
52 On/off config. O
2
Auto range (Fct.4O)
Store function
(Fct.77)
...continue
Figure 5-36 On/off functions
With this function, you can easily switch other functions on/off, such as the ones listed in the
above function screen.
Thanks to this simplified operation, these functions do not have the longer paths through
various menus. For better orientation, the function numbers have also been specified.
Activated functions are marked by "■", deactivated ones by "□". Using the fifth softkey
("...continue"), jump to the next function screen with further functions.
Function 52 is used to switch the following functions on and off:
Table 5- 3 Functions switched on/off by function 52
Description Fct. No. Remarks
Automatic measuring range switching 40
Measured-value memory 77
Temperature recompensation of zero point 86
Temperature recompensation of span 86
Signal tolerance violation 78
Limit monitoring 1 51
Limit monitoring 2 51
Limit monitoring 3 51
Limit monitoring 4 51
Suppress negative measured values 70
Automatic calibration 24 Only with supplementary
electronics
Total calibration 23
Block logbook 60
Fault / WA / CTRL MAMUR 72
Suppress negative measured values for display 55
Besides the functions listed in the table "Functions switched on/off by function 52", function
52 can also be used to address other service functions. These are restricted to service
engineers and are only visible when the service code is entered (code level 3).