Installing and connecting
3.4 Electric connection
OXYMAT 64
Operating Instructions, 06/2008, A5E00880383-03
29
The RC element should therefore be dimensioned according to the following rule of
thumb:
R [ Ω ] ≈ 0.2 x R
L
[ Ω ]
C [μF] ≈ I
L
[ A ].
Also make sure that a non-polarized capacitor C is used.
In the case of operation with direct current, a spark-quenching diode can be fitted instead
of the RC element.
The connection lines must be shielded to the relay outputs, the binary inputs, the analog
inputs and the analog output. Connect these to the corresponding trapezoidal plugs
(SUB-D plug connectors) in accordance with the routing assignments.
68%')FRQQHFWRU
3RZHUVXSSO\9PD[

























5
&
,/
5/
5>˖@[5
/
>˖@
&>˩)@,
/
>$@
 
0
0
Figure 3-3 Example of spark suppression on a relay contact
The reference ground of the analog inputs is the potential of the housing.
The analog output is floating.
The shielding of the interface line must be at the potential of the housing. Connect the
shield of the line to the shield of the SUB-D plug connector with a high contact area. The
core cross-section must be ≥ 0.5 mm². The interface line may be at most 500 m long.
Note
All connection lines (except for the mains connection line) must be shielded. The
shielding of the connected lines must have a large-area and continuous contact at the
respective screw couplings. The core cross-section must be ≥ 0.5 mm². We recommend
cables of type JE-LiYCY...BD. The cable length of the analog output depends on the
load.