Operation
5.4 Operating functions
OXYMAT 64
Operating Instructions, 06/2008, A5E00880383-03
83
The connection assignments for the individual relays in the zero-current state can be found
in the routing assignments "Motherboard pin assignments". The relays are preset at delivery.
Press the fifth softkey ("...continue"), to access the next function screen, thus branching to
further relays.
Table 5- 6 Relay outputs (possible functions)
Function Relay is passive
when
Relay conducts current Remark
Not assigned Relay is permanently
passive (zero current)
Fault Fault Also shown in the status
line of the measured
value display
Maintenance request Maintenance request
Calibration Calibration running For identification
purposes
Measuring range 1 to 4 Measuring range 1 to 4
on
Measuring range
identification
Limit 1 to 4 Limit 1 (to 4) triggered Limit monitoring
Functional check
(CTRL)
Functional check
active (e.g.: when
changing to "Operator
control mode", in the
warm-up phase, when
Autocal is running,
when calibration is
running)
Functional check
deactivated
Signal when:
device is decoded,
warm-up phase (max. 2
h), calibration running
(Autocal)
Sample gas* Supply of sample gas Valve actuation
Zero gas 1 to 2* Supply of zero gas
Calibration gas 1 to 4* Supply of calibration gas
Measuring point 1 to 6 Measuring point 1 to 6
selected
For taking a gas sample
via solenoid valves at
various measuring points
Measuring point signal
1 to 6
Measuring point 1 to 6
selected
For identifying the
measuring point (runs
parallel to measuring
point)
Signaling contact* When there's a signal,
the relay conducts
current for a short time
e.g. in the case of
AUTOCAL:
actuation of a second
device
Autocal check Autocal difference too
large (function 24)
Dimension % v/v vpm % For identification
purposes
* This function is only available if the analyzer is fitted with an optional Autocal board
See also
Motherboard pin assignments (Page 30)