Service Manual SC 7000 and SC 9000XL Patient Monitors
142 Siemens Medical Systems, EM-PCS, Danvers ASK-T898-03-7600
NOT A CONTROLLED DOCUMENT
7k9kXLSM.ape.CD_ROM.fm/04-99/kaupp
16. Central Station enabled: yes or no: If you choose yes, bedside monitor
expects an alarm acknowledgment from MVWS. If it doesn’t get one, it
goes to its highest alarm volume and indicates an off-line condition.
If you choose no, bedside monitor does not expect an acknowledg-
ment from MVWS and will not go to its highest alarm volume.
If there is no M
ULTI
V
IEW
W
ORK
S
TATION
on this I
NFINITY
N
ETWORK
, or
this bed is NOT to be centrally monitored, choose “NO.”
4 Configuring Monitor
1. On Main menu, access Monitor Setup Unit Manager, and enter
Clinical password (
375
).
2. Select Menu Setup, and set Menu Time Limit to OFF.
3. On Main menu, access Monitor Setup Biomed Service, and enter
Biomed password (
4712
).
4. For each configuration parameter, enter data from Table 1 and select
“Accept”.
Note: For numerical fields, rotary knob increments and
decrements numbers in the field as well as enters the data.
5. After all data has been entered, recheck data and then select Save
ALL.
Note: Monitor saves all entries, and then power-cycles if
Network Mode or any IP address parameters were changed.
Menu Time Limit automatically resets to ON if monitor
power-cycles.
6. Power-cycle monitor, unless monitor power-cycled in 5.
7. Connect monitor to network, and verify on MVWS that monitor
communicates with MVWS and configuration contains no duplicate
names.
8. This completes monitor setup.
Table 4 Monitor Configuration Parameters
Network Config
Network Mode
Direct Net.
Host ID
z
Bed Monitoring Unit ID
y
CPS/IDS IP Address
191.1.
y
.
z
Care Unit Subnet Mask
255.255.0.0
Monitoring Unit Default Route
0.0.0.0
Hospital Remote Silence
Recorder 1
Remote Control
Recorder 2
Alarm Group
Recorder Use Network Central Station
Save ALL