Service Manual SC 7000 and SC 9000XL Patient Monitors
72 Siemens Medical Systems, EM-PCS, Danvers ASK-T898-03-7600
NOT A CONTROLLED DOCUMENT
7k9kXLSM.c4.CD_ROM.fm/04-99/kaupp
3) Press C.O. START key on simulator, and verify an Injectate
Temperature indication of 1±0.2°C.
14Memory Backup
Function
The monitor retains patient-related data, such as alarm limits, trends, and
stored alarm recordings when it is powered off.
1) With monitor powered-up and no patient inputs applied, press Alarms
Limits fixed key and change limit for any parameter.
Note: Note time on Real Time Clock.
2) Power monitor off for approximately 2 minutes, and then power it back
on again.
3) After
MAIN
screen displays, press Alarm Limits fixed key to call up
Alarm Limits Table.
4) Verify that new limit(s) you set in Step 1 have been retained, and that
clock has advanced the proper amount of time.
15CPS/IDS Mode
This procedure tests that monitor communicates with a CPS or IDS. Omit
this test and go on to Section 16 if monitor used exclusively in DirectNet
mode.
1) With monitor mounted on docking station connected to a CPS, or on
IDS press Menu key, and select Monitor Setup Biomed Service.
2) Enter biomed password (
4712
) and select “Accept.”
3) Select Network Setup Network Config. CPS/IDS Save ALL.
Note: Monitor resets if Network Config. previously set for DirectNet.
4) Press Menu key and select Monitor Setup Biomed Logs
5) Select Component Log CPS/IDS.
6) Verify that Serial # of CPS or IDS appears on display.
16DirectNet Mode
(requires
VC2-level
installed software)
This checks for proper configuration of a monitor set up for DirectNet
functioning, and for monitor communication via the I
NFINITY
N
ETWORK
. Omit
this test if monitor used exclusively in CPS/IDS mode.
1) With monitor disconnected from docking station, connect monitor to
I
NFINITY
N
ETWORK
using ethernet connector on right side of monitor.
Note: Remove label covering ethernet connector on right-hand side of
monitor, if necessary.
2) Press Menu key, and select Monitor Setup Biomed Service.
3) Enter biomed password (
4712
) and select “Accept.”
4) Select Network Setup Network Config. DirectNet Save ALL.
Note: Monitor resets if Network Config. previously set for CPS/IDS.
5) Setup monitor for simulated patent using external patient simulator.
6) Verify that waveforms display on MVWS.
Note: If I
NFINITY
N
ETWORK
not equipped with MVWS, use remote view
function of another monitor on network to view waveforms. Then skip
step 7 and go to step 8.
7) Access
netInfo
on MVWS and verify that monitor labels are unique.
8) Follow procedure in steps 2 through 4 to reset monitor to CPS/IDS
mode, if required.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SC 7000 and is the answer not in the manual?

Siemens SC 7000 Specifications

General IconGeneral
BrandSiemens
ModelSC 7000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals