Connecting
3.2 Sensor Wiring
FUT1010 IP65 NEMA 4X & IP66 NEMA 7 Liquid Quick Start
Operating Instructions, 1/2013, A5E02639182-AC
31
1. Remove four (4) #10 bolts securing Junction Box top cover. Remove cover and set aside
bolts.
83
83'1 '1
&+$1 1(/
&+$1 1(/
&+$11(/ &+$11(/
83 83'1
'1
7%









CH 4-Sensor to Junction Box Gland CH 4 - UP to Transmitter
CH 3-Sensor to Junction Box Gland CH 2 - DN to Transmitter
Top Cover #10 bolts (4) CH 2 - UP to Transmitter
Temperature Sensor Terminal Board
TB1
Temperature Sensor Gland
CH 2-Sensor to Junction Box Gland CH 1 - DN to Transmitter
Transducer Channel F-Connectors CH 1 - UP to Transmitter
Junction Box without top cover CH 3 - DN to Transmitter
CH4 - DN to Transmitter CH 3 UP to Transmitter
CH 1-Sensor to Junction Box Gland
Figure 3-6 Sensor Cable Installation
2. Observing the upstream to downstream orientation, pull Channel-1 and Channel-2 UP
and DN Sensor cables into the corresponding CH-1 and CH-2 Junction Box glands.
NOTICE
Incorrect Bend Radius
If bend radius is incorrect d
amage to cables may result.
When connecting sensor cables inside Junction Box cable bend radius should not
exceed bend tighter than 8 cm (3 in.).