3.2 Sensor Wiring
FUT1010 IP65 NEMA 4X & IP66 NEMA 7 Liquid Quick Start
Operating Instructions, 1/2013, A5E02639182-AC
33
3.2.2 Wiring Temperature Cable to Sensor
1. Locate Temperature Sensor terminal block TB1 connection screws.
2. Insert Temperature Sensor cable from transmitter into Junction Box gland.
7%
776
7%
775
83
83'1 '1
&+$11(/
&+$11(/
&+$11(/ &+$11(/
83 83'1
'1
7%

Black Short Terminals 1 and 4
Orange Ground Terminals 2 and 3 to Terminal 5
Brown 1012EC Series Cable
Red Junction Box with cover removed
Blue Temperature Sensor Connections: TB1-1
Black/Orange and TB1-2-Brown/Red. Tie off Blue
Figure 3-8 Temperature Sensor Board to Junction Box Wiring
3. Connect Temperature Sensor PCB wires to Junction Box Sensor terminal board as
follows:
Insert Black and Orange wires into lug (supplied), crimp lug and connect lug to TB1-1.
Insert Brown and Red wires into lug (supplied), crimp lug and connect lug to TB1-2.
Tie off Blue wire.
Make sure all terminal block lug connections are hand-tight.
4. Replace Junction Box top cover.
5. Reinstall and hand-tighten four (4) Junction Box securing bolts. Using a torque wrench,
torque each bolt to 6.8 to 8.1 Nm (5 to 6 ft-lbs).