Additional Installation Instructions
B.4 Hi Precision mount
FSS200 clamp-on sensors
116 Installation Manual, 08/2017, A5E36255466-AC
3. Place the enclosure housing on the pipe and mark pipe at each strap location using chalk
or a grease pencil.
4. Prepare straps for mounting by cutting to required length. Strap length should be twice
the pipe circumference plus 15.24 cm (6 inches).
5. Remove set screw from strap fastener. Retain set screw.
6. After strap is cut to correct length feed strap into strap fastener (small set-screw clamp).
Bend 2.54 cm (1-inch) of the end lip around the strap fastener.
Figure B-27 Bending strap end
Note
For Reflect mount, install strap retainer for each path (i.e. Single path Reflect mount will
have one strap retainer, Dual path Reflect mount will have two strap retainers at each
strap location). For Dire
ct mount, install two strap retainers at each strap location.
7. Mount the straps at the previously marked locations. Loop strap around pipe and feed the
open end of the strap through the strap fastener and strap retainer. Pull loose end of
strap to remove slack.