7 SERVICE
100
SITRANS FX330
www.siemens.com/flow 08/2022 - A5E36751269-AG EN
The following steps are to be performed:
• Switch off the power supply.
• Unscrew the front cover with the key.
• Use the handle to pull out the display by plugging into the brackets ④ and gently pulling it out.
• Detach the flow sensor connection ⑤ .
• Loosen the two fixing screws ⑥ .
• Pull out the signal converter.
• Insert the new signal converter.
• Tighten the two screws ⑥ .
• Attach the flow sensor cable ⑤ .
• Attach display ① in desired position,
apply even pressure to the entire surface.
• Screw on cover by hand.
INFORMATION!
Use only spare parts that are unused and do not yet contain any configuration data.
Never connect components that contain data of two different devices!