EasyManuals Logo

Sigicom INFRA C22 User Manual

Sigicom INFRA C22
77 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #68 background imageLoading...
Page #68 background image
Copyright © Sigicom AB 2021
68
70A. BS 6841
Acceleration standard including channels for VDV (Vibration Dose
Value).
See Appendix H for further description.
Level and frequency ranges: 125 m/s
2
.
Max resolution in presentation: 10
-4
m/s
1.75
and 3 significant digits.
Threshold range: 0.1 – 110 m/s
1.75
.
70B. BS 6841
Acceleration standard including channels for VDV (Vibration Dose
Value).
See Appendix H for further description.
Level and frequency ranges: 12.5 m/s
2
.
Max resolution in presentation: 10
-4
m/s
1.75
and 3 significant digits.
Threshold range: 0.04 – 11 m/s
1.75
.
71A. BS 7385 & 6841(VDV)
Evaluation and measurement for vibration in buildings - Part 2: Guide to
damage levels from ground-borne vibration. In addition the interval
channels VDV and VDV-accu are calculated (see Appendix H).
Level and frequency ranges: 250 mm/s, 1-300 Hz.
Max resolution in presentation: 0.05 mm/s and 3 significant digits.
Threshold range: 0.4 220 mm/s.
71B. BS 7385 & 6841(VDV)
Evaluation and measurement for vibration in buildings - Part 2: Guide to
damage levels from ground-borne vibration. In addition the interval
channels VDV and VDV-accu are calculated (see Appendix H).
Level and frequency ranges: 25 mm/s, 1-300 Hz.
Max resolution in presentation: 0.005 mm/s and 3 significant digits.
Threshold range: 0.04 22 mm/s.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sigicom INFRA C22 and is the answer not in the manual?

Sigicom INFRA C22 Specifications

General IconGeneral
BrandSigicom
ModelINFRA C22
CategoryMeasuring Instruments
LanguageEnglish