Example 1 (Drywall):
Suppose that selected trigger level is 100%. Then a transient will be
triggered when for example:
a signal (peak) is greater than 0.38 in/s and its frequency is 2 Hz
a signal (peak) is greater than 0.75 in/s and its frequency is in the
range 4 - 15 Hz
a signal (peak) is greater than 2.0 in/s and its frequency is 40 Hz
(or higher)
Example 2 (Drywall):
Suppose that selected trigger level is 50%. Then a transient will be
triggered when for example:
a signal (peak) is greater than 0.19 in/s and its frequency is 2 Hz
a signal (peak) is greater than 0.38 in/s and its frequency is in the
range 4 - 15 Hz
a signal (peak) is greater than 1.0 in/s and its frequency is 40 Hz
(or higher)
Example 3 (Plaster):
Suppose that selected trigger level is 100%. Then a transient will be
triggered when for example:
a signal (peak) is greater than 0.38 in/s and its frequency is 2 Hz
a signal (peak) is greater than 0.5 in/s and its frequency is in the
range 2.6 - 10 Hz
a signal (peak) is greater than 2.0 in/s and its frequency is 40 Hz
(or higher)
Example 4 (Plaster):
Suppose that selected trigger level is 200%. Then a transient will be
triggered when for example:
a signal (peak) is greater than 0.76 in/s and its frequency is 2 Hz
a signal (peak) is greater than 1.0 in/s and its frequency is in the
range 2.6 - 10 Hz
a signal (peak) is greater than 4.0 in/s and its frequency is 40 Hz
(or higher)