Instruction Manual LabScat 2 Servicing
44 12112E/2
WORKSTEP
ADDITIONAL INFO / IMAGES
8.
Perform adjustment.
Press the initiate button and wait.
If the adjustment was successful, confirm with
Adjustment OK. This completes the adjust-
ment.
If the adjustment was not successful, it is indi-
cated with Adjustment fault. In this case,
check the points in the following list one after
the other:
Cleanliness of the checking unit.
Correct checking unit used.
Nominal values correspond to the values of
the checking unit.
If the check was not successfully com-
pleted, contact your country representative.
With the initiation of the 90°
adjustment, the 25° adjustment is
also always initiated.
9.
Remove the checking unit.
10.
The instrument can now be operated again.
A new recalibration factor is determined during recalibration. The deviation from the original
state is displayed under Curr. corr..
The text file recali.txt is also created that can be viewed using a text editor. This is copied to
the USB stick in the SIGRIST_LOG folder.