5 Care, cleaning and maintenance by the practice team Sirona Dental Systems GmbH
5.4 Vacuum system Operating Instructions SINIUS / SINIUS CS
Amalgam residues, amal gam rotor and sedi ment container
The outside of the suction hoses can be disinfected by spraying or wiping.
Assistant element, suction hoses
Thermodisinfectable hoses are available as special equipment; see
Thermodisinfection of suction hoses [ 197].
Wear gloves, risk of contamination
1. Switch the treatment center off at the main switch.
2. Pull the filter housings (B) off of the connectors on the assistant
element.
3. Disconnect the suction hoses from the filter housings (B).
4. Take the collectors (C) out of the suction hoses. Collect the amalgam
residues in a glass filled with water.
5. Take off the suction cannulae and rinse out the suction hoses with
clean water.
If the treatment center is equipped with thermodisinfectable suction
hoses, thermal disinfection can be performed after cleaning, see
"Thermodisinfecting the suction hoses" [ 197].
Assembly is performed in reverse order. Grease the O-rings (A) before
reconnecting the suction hoses. For greases, see "Care and cleaning
agents" [ 170]. Make sure that the suction hoses snap into place.
CAUTION
Amalgam residues must not enter the public sewage system.
Amalgam is a mercury compound that is hazardous to water.
Do not dispose of amalgam residues in a sink.
Collect amalgam residues e.g. in a closed container with water.
Dispose of amalgam residues, e.g. when replacing the amalgam
rotor by filling the amalgam residues into the amalgam rotor or when
emptying the sediment container.
NOTICE
Powdering suction hoses with talcum
If the surfaces of the hoses have become sticky from disinfectants,
clean them with a commercially available dishwashing liquid and then
powder them with a light coat of talcum if necessary.
!