67
Sirona Dental Systems GmbH 4 Operation
Operating Instructions SINIUS / SINIUS CS 4.4 Patient chair
båÖäáëÜ
4.4.7 Moving the patient chair via chair programs
Moving the patient chair via chair programs
Chair programs 0 and S on the assistant element, references to “Creating chair p rograms”
The chair programs can be selected via the touchscreen. The entry/exit
and mouth rinsing positions can also be selected via the assistant
element fixed keys and the foot control.
You can individually re-program the factory preset chair programs to suit
your own requirements, see "Creating chair programs and shock
positioning" [ → 72].
Swivel arm dentist element, collision, program run
Cuspidor, manual, starting position
4.4.7.1
Moving the patient chair to the entry/exit position
Moving the patient chair to the entry/exit position
The following functions are triggered for simple patient entry and exit in
the entry/exit position:
Entry/exit position, without motor-driven travel track and without motor-driven cuspidor
● The patient chair moves to an upright position
● The operating light switches off
The tumbler heater can be set so that it automatically switches off when
the entry/exit position (0) chair program is activated; see "Linking the
tumbler heater to the entry/exit position" [ → 163].
Dentist element, hoses, entry/exit
Via touchscreen
✔ The
Start program
is displayed on the touchscreen.
➢ Touch the
0
key briefly (< 2 s).
WARNING
The SINIUS CS dentist element could be positioned within the
movement range of the patient chair.
During an automatic program run, such as approaching the entry/exit or
mouth rinsing positions, the patient may collide with the SINIUS CS
dentist element or its support arm. The patient is at risk of being
crushed.
➢ Before moving the patient chair, position the SINIUS CS dentist
element in such a way that a collision with the patient or the patient
chair is impossible.
IMPORTANT
Chair movements with cuspidor swiveled in
With the cuspidor bowl swiveled in, the chair cannot be moved. This
prevents the patient from colliding with the cuspidor. Swing the cuspidor
outward before initiating chair movement.
WARNING
The patient's feet may get caught in the instrument hoses of the dentist
element when he enters or leaves the patient chair.
The patient may trip or fall.
➢ Turn the dentist element outward before the patient enters or leaves
it.