Sirona Dental Systems GmbH 4 Operation
Operating Instructions SINIUS / SINIUS CS 4.5 Dentist element
båÖäáëÜ
1. Attach the NaCl bottle holder to the dentist element.
2. Hang the NaCl bottle (max. 1 liter) onto the bottle holder.
3. Undo the locking mechanism A of the peristaltic pump and fold it
down. Swivel the latch fastener (B) upward.
4. Open the two hose holders (C) upward (at the top) and downward.
Lay the silicone hose (D) without prestress, i.e. with the thickened
part wrapped around the pump wheel. See the adjacent drawing.
Close the hose holders (C).
5. Swivel the latch fastener (B) down onto the hose on the pump wheel.
Secure the latch fastener B with the locking mechanism A.
IMPORTANT
Direction of flow of the peristaltic pump
The shorter end of the hose with the cannula must be located at the top
of the pump; the longer end of the hose leading to the handpiece must
be located at the bottom. See the adjacent drawing.
IMPORTANT
Flow rate of the peristaltic pump
The flow rate is set with the spring on the locking mechanism (A). Do not
under any circumstances change the factory setting.