Maintenance, Troubleshooting, Transport and Storage
10-2
Table 10-1 Troubleshooting List
Failures Causes and Solutions
With power supply, but no
image on the monitor
The monitor brightness and contrast are adjusted too low,
so the images cannot be seen. Please adjust the
brightness controls and the contrast controls to
appropriate positions.
No ultrasonic image
displayed on screen in any
scanning mode, but the
characters and grayscale bar
are displayed normally
Inspect if the probe is correctly connected to the main
unit.
The printer not working
Check if the printer is correctly connected to the main
unit, and if its power is on;
If there is no reaction by pressing
Print key on the
keyboard, but the
Print key on the printer responds by
pressing, then it is possible that the print remote-control
cable is not well connected.
Incorrect date display Reset the date referring to Section 5.2.2.1.
If any abnormality occurs, please immediately turn off the power, and inform our nearest
representative about the abnormal circumstance as detailed as possible.
Please contact us for help whenever it is necessary to repair or readjust the equipment.
We shall not be responsible for any failure caused after other people’s repairing or
readjusting.
Note: In case any abnormal situation occurs on the system, to enable good service
within a short period, please try to describe the problem as detailed as
possible and send it to us.
Statement: Circuit schematics and component list may be provided to SIUI
approved qualified service personnel for equipment maintenance.
WARNING: No modification of this equipment is allowed.