Application Scope, Operation Conditions, Safety and Cautions
1-1
Chapter 1
Application Scope, Operation Conditions,
Safety and Cautions
1.1 Application Scope
The system is a general-purpose diagnostic ultrasound imaging system designed exclusively
for use in a wide variety of extracorporeal body imaging procedures. A general-purpose
system supports various transducers and related application software packages allowing for
the collection, display and analysis of ultrasound information. Usages include, e.g.
general-purpose imaging, cardiac, OB/GYN, breast, prostate, vascular, intra-surgical,
depending on the operating system specific software packages and compatible ultrasound
transducers (this product does not support intra-surgical and endocavity transducers).
1.2 Operation Conditions
a) This system should be operated under the following ambient conditions for the safe and
correct operation:
z Ambient Temperature: 0 to 40
z Relative Humidity: 30% to 85%
z Atmospheric Pressure: 700hPa to 1060hPa.
Note 1: If the operation condition is beyond the ranges above, no correct ultrasonic
images are ensured.
Note 2: If the system is stored for a long time, the ventilation of the stored place shall be
considered.
Note 3: The system is air cooled with a fan.
b) Strong sources of radiated emissions or electromagnetic waves, from broadcast & TV
station for instance, may cause the system to operate incorrectly or to display with noise.