Application Scope, Operation Conditions, Safety and Cautions
1-2
The system must be separated from such emission sources or electromagnetic waves.
c) Do not operate this system while other devices are operating with motor or silicon
controlled switch in the same power phase; otherwise, noise will disturb your system
through power cable.
1.3 Power Requirements
Never use this system when the power supply does not meet the requirements specified
below, otherwise, the system may be damaged.
a) Main Unit Power Voltage:
Battery: DC12V-16.8V, or
External Power Adapter: DC12V±1.2V
b) Accessories Power Voltage:
Adapter External Power Supply: 100V-240V~;
Battery Charger External Power Supply: DC12V±1.2V.
c) Power Frequency: 50Hz±1Hz or 60Hz±1Hz for power adapter.
d) If the non-medical electrical equipment, supplied as a part of the system, is intended to
be supplied via a multiple portable socket-outlet with a separating transformer, the direct
connection of the non-medical electrical equipment to the wall outlet may result in risks;
e) Do not connect any additional removable multiple socket or cables to the removable
multiple socket used by the system devices or to the system devices;
f) Do not connect any accessories or devices that are not described in this manual to the
unit or the system built up as per this manual. Otherwise it would lead to unexpected
danger;
g) When the system built up as per this manual is in use, all the device power plugs shall
be connected to the same special removable multiple socket. If the removable multiple
socket is supplied by a special power device, the power device shall comply with
IEC60601-1 and IEC60601-1-1 requirements. The removable multiple socket used for
the system shall comply with requirements in Annex EEE of IEC60601-1-1.
h) Do not place the removable multiple socket on the ground, otherwise, it may result in
danger;
i) The maximum load capacity of the removable multiple socket shall not be less than the