Acoustic Output Parameter
D-7
Transducer Model:U5I50H Operating Mode:M
TIS TIB
Non-scan
Index Label
MI
Scan
A
aprt≤1cm² Aaprt>1cm²
Non-
scan
TIC
Global Maximum index value 0.911 0.097 0.141 0.106
Prα Mpa 2.052
P mW 4.0 4.0 4.0
Min.of[Pα(zs),Ita,α(zs)] mW
zs cm
zbp cm
zb cm 1.0
z at max. Ipi,α cm 2.80
deq(zb) cm 0.462
fawf MHz 5.078 5.078 5.078 5.078
X cm 1.17 1.17
Associated
Acoustic
parameters
Dim of
A
aprt
Y cm 0.60 0.60
td μsec 0.455
p
rr
Hz 515
pr at max. Ipi Mpa 3.29
deq at max. Ipi cm 0.259
Other
information
I
pi,α at max.MI W/cm² 242.88
Control 1:
Power
dB 0 0 0 0
Control 2:
Focal Depth
cm 4.0 4.0 4.0 4.0
Control 3:
M Speed
s 1.25 1.25 1.25 1.25
Operating
Control
conditions
Control 4:
Frequency
MHz 5.0 5.0 5.0 5.0