Use and Operation
6-8
Menu No. Menu Menu Function
factory
defaults
correct the
filament value
window; and the mAs window displays
the current corrected Kv/mA parameter
combination in turn.
P1 Reset
Restore to factory settings, including the
imaging parameters of APR, tube
filament value, and feedback control
time of kV. The device is shut down after
reset.
N/A
P3
Exposure
number
To calculate the exposure number upon
each exposure. Press SAVE key in the
menu to clear the exposure number.
Not cleared
P4
Program
version
Display the current system software
version number, as well as power and
device version..
N/A
P6 Standby time
The default value is 600 seconds. The
value can be adjusted in the range 【30,
1800】at the increment of 30s.
When
adjusted to 0, there is no standby time
limit.
600 (seconds)
P10
Collimator
light up time
The default value is 120 seconds. The
value can be adjusted in the range 【30,
300】at the increment of 30s.
120 (seconds)
P15
Control
program
update
Program update through the USB disk.
N/A
P21
The timing
setting from
Press the shutdown button, the timing of
the automatic power down will
30 (minutes)