Use and Operation
6-9
Menu No. Menu Menu Function
factory
defaults
shutdown to
poweroff
operation. The default value is 30
minutes. The value can be adjusted in
the range 【0, 120】at the increment of 1
minute. When it is adjusted to 0, the
timing limit is not performed.
P22 Wi-Fi function
0-Close Wi-Fi function
1- Open the Wi-Fi function
0 (close Wi-Fi
function)
P26
Exposure
mode setting
Level2 exposure mode
1- Level1 exposure mode
exposure
mode)
P27
time setting
of one-press
exposure
button
The default is 5 (seconds).
The setting range is in the range 【2,
15】at the increment of 1 (second)
5 (seconds)
P28
Exposure
delay setting
of one-press
exposure
button
The default is 5 seconds.
The setting range is【0, 60】at the
increment of 1 (second). When it is
adjusted to 0, no delay is performed and
the exposure is performed immediately.
5 (seconds)
P29
View the
number of
exposure for
high mAs
Count the number of consecutive
exposures of mAs ≥ 30 mAs. When the
parameter save button is pressed in this
menu, the number of exposures is not
cleared.
Not cleared
P33
exposure
mode setting
Set 0 to one-key exposure OFF set 1 to
one-key exposure ON
exposure
OFF)
P36
Modify the
remote IP
address
It is valid when P40 is 0. APR1-5 is to
select 5 different remote devices.
Device 0:
192.168.8.188
Device 1: