Set
PARAMETER
DESCRIPTION
POSSIBLE SUPERUSER ACTION
With set parameter
-d=0”
With set parameter
-d=0”
-9
Cycles left before maintenance
Indicates, to the nearest ten, the number of cycles left
before the maintenance message appears. The initial
default setting is 900 (e.g. “-9:90”).
The value can be reset by an authorised engineer.
Display Only
Display Only
-A
Power Supply
The parameter indicates whether the connection is:
- single-phase: parameter setting “1”
- three-phase: parameter setting “3”
N.B.: the value “3” refers to both 400V 3N~ / 50 Hz and
230V 3~ / 50Hz connections.
Warning: this parameter is only used for the correct
setting of times remaining.
Display Only
Display Only
-b
Memory Data Overwriting
- Overwriting enabled: parameter setting “1”
- Overwriting disabled: parameter setting “0”
The parameter refers to whether or not the device’s
internal memory, which keeps a record of all the cycles
performed and the alarms which have occurred, can be
overwritten.
WARNING
If overwriting is disabled, alarm AF:91 is triggered once
the memory is full. In this case the appliance will not
perform any more cycles until the data from the
memory are downloaded; this can only be done using
the TRACELOG software.
Display and
Modification
Display and
Modification
-d
Superuser dosage setting disabling parameter.
(Parameter introduced from Main Firmware version
4.6.xx.yy).
If “-d: 0” the superuser is able to modify the dosages
If “-d: 1” the superuser is not able to modify the dosages
but can display them.
Default value: “-d: 0”.
This Parameter can only be modified by the authorised
engineer.
Display Only
Display Only
-E
Parameter for setting presence or absence of the WD-
LANE accessory.
(Parameter introduced from Main Firmware version
4.6.xx.yy).
Sets the serial port communication speed
-E: Pr” = 9600 bps (for Printer connection)
-E: Ln” = 11500 bps (for connection to WD-LANE - LAN
accessory).
Display and
Modification
Display and
Modification