Tethys D60 - Tethys T60 - Tethys T45
6.3 “SET“ PARAMETERS
Set on the display.
To access this item press the Start/Stop button.
The Set item contains a number of setting parameters.
User level operators can only display the parameters and cannot modify any “Set” parameters.
POSSIBLE SUPERUSER ACTION
With set parameter
“-d=0”
With set parameter
“-d=0”
Detergent pump P1 dispensing rate.
Setting in ml/litre, with maximum threshold of 20
ml/litre.
The default value of this parameter is 4ml/litre.
Maximum recommended quantity with recommended
detergents: 10ml/litre.
Increase and decrease with the Selection buttons.
Detergent pump P2 dispensing rate.
Setting in ml/litre, with maximum threshold of 20
ml/litre.
The default value of this parameter is 2ml/litre.
Maximum recommended quantity with recommended
detergents: 10ml/litre.
Increase and decrease with the Selection buttons.
Detergent pump P3 dispensing rate.
Setting in ml/litre, with maximum threshold of 20
ml/litre.
The default value of this parameter is 10ml/litre.
Maximum recommended quantity with recommended
detergents: 15ml/litre.
Increase and decrease with the Selection buttons.
Water hardness setting
Set the hardness in °f, in steps of 5. Range: 5-60°f. The
default setting is 40°f.
Warning: Make sure that the value is actually the same
as the hardness of the water used.
Printer language
English by default, the language can be chosen from the
various options available
Demineralised water
The use of demineralised water can be activated or
deactivated depending on whether or not the relative
connection has been made (Activated: dn, Deactivated: -
-)
Filter hours (Tethys D60 only)
The parameter indicates, to the nearest ten, the number
of absolute filter hours left before the maintenance
message appears. The initial default setting is 500. (e.g.
the display shows “-8:50”). The value can be reset by an
authorised engineer.