1 INTRODUCTION
Tethys D60, Tethys T60, Tethys T45 - instrument washers for use in hospitals, class IIb medical devices (in accordance
with the classification criteria established by Directive 93/42 and subsequent additions and amendments, extended
and amended by EC Directive 2007/47).
Tethys family products comply with all the prerequisites of the relevant safety regulations and the standards in force
for electrical equipment.
This manual describes products series.
• Tethys D60: instrument washers, 60cm, equipped with forced air drying system.
Tethys T60: instrument washers, 60cm, without drying system.
• Tethys T45: instrument washers, 45cm, without drying system.
The diagrams and images contained in this document refer to the Tethys D60 series. For the Tethys
T60, Tethys T45 series the same considerations and procedures apply, unless otherwise specified.
The table below summarises the product’s main characteristics. (For a full list of models and their configurations refer
to the “LIST OF MODELS” point below).
Basic models
CONTROL
ELECTRONIC with MICROPROCESSOR
(with Time and Date functions)
DOOR OPENING
Automatic
DETERGENT DISPENSING
Automatic, by means of
Peristaltic pumps (P1 and P2)
OPTIONAL DETERGENT DISPENSING
Optional peristaltic pump P3
CONTROL OF DETERGENT DISPENSING
Yes
DETERGENT JERRY CAN LEVEL CONTROL
Yes (Tethys D60)
TEMPERATURE CONTROL
Two probes, work and monitoring (TL1 and TCL)
STEAM CONDENSER
Yes
DRYING SYSTEM
Yes, forced air (Tethys D60 series Only)
ABSOLUTE DRYING FILTER
Yes (Tethys D60 series Only)
WATER INTAKE CONNECTIONS
Mains and Demineralised
PRESET PROGRAMS
Specific HOSPITAL, TOTAL OF 16 PROGRAMS, 6
“CUSTOM” (customisable using
TRACELOG software
1
. The software is not included in the
product’s basic outfit)
COMMUNICATION
RS-232 serial (For PC or Printer)
1
The TRACELOG software enables the user to communicate with Tethys instrument washers.
Once the connection has been made, the user can:
- Keep a record of the disinfection cycles performed.
- Monitor the appliance’s status.
- Download new versions of the appliance’s software.
- Modify programs.
- Save the appliance parameter setup.