Tethys D60 - Tethys T60 - Tethys T45
CHARACTERISTICS
RECOMMENDED
MAGNETOTHERMAL
SWITCH
[IN THE SYSTEM, RESPONSIBILITY
OF THE USER]
Max voltage/power
There are three possible
electrical versions.
60cm models: The names of the
models have different suffixes
to identify the power supply
rating.
E.g. If the model does not have
a suffix, e.g. “Tethys D60” it will
be three-phase, with neutral,
400V between phases.
“Tethys D60-3” indicates three-
phase model, 230V between
phases, no neutral.
“Tethys D60-1” indicates single-
phase model.
400V 3N~ / 50Hz / 12A / 7000W
FUSES 10.3x38 16 A ON
POWER SUPPLY
[L1, L2, L3]
Tethys D60-3
Tethys T60-3
230V 3~ / 50Hz / 19A / 7000W
FUSES 10.3x38 20 A ON
POWER SUPPLY
[L1, L2, L3]
Tethys D60-1
Tethys T60-1
230V 1N~ / 50Hz / 12A / 2800W
FUSES 10.3x38 16 A ON
POWER SUPPLY
[L1, N]
230V 1N~ / 50Hz / 14A / 3300W
FUSES 10.3x38 16 A ON
POWER SUPPLY
[L1, N]
DATA COMMUNICATION
- PRINTER CONNECTION
80% for temperatures up to 31°C with linear decrease
to 50% at the temperature of 40°C
INSTALLATION CATEGORY (OR SURGE CATEGORY or
INSTALLATION CATEGORY)
ELECTRICAL INSULATION CLASS (ref. IEC 61140)
2 (rif. IEC61010-1:2010, par.3.6.8)
STANDARDS AND CLASSIFICATION
- European directive 2014/35/EU (2006/95/CE) [Safety]: European Standard EN61010-
1:2010, EN61010-2-040:2015
- European directive 2014/30/EU (2004/108/CE) [Emc] : European Standard
EN61326:2013
- European directive 2011/65/CE (RoHS 2) : European Standard EN50581:2012
- European directive 93/42/CE, 2007/47/CE [MD]: European Standard
EN ISO 14971:2012, EN 62304:2006+AC:2008, EN62366:2008, EN ISO 15883-
1:2009+A1:2014, EN ISO 15883-2:2009
- Other: European Standard EN 61770:2009
MEDICAL DEVICE
CLASSIFICATION
class IIb medical devices (in accordance with the classification criteria established by
Directive 93/42 and subsequent additions and amendments, extended and amended by
EC Directive 2007/47).