wi lapici o
{RECALL}
> >
Remote
command is RC
I. Refer to Chapter
4,
Section 6.25.
Enter an integer
’n’ to recall the
contents of store
’n’ to set
up the stored menu
selections.
The sequence
is then similiar
to that for
STORE, except
that ENTERing
the
required
store number,
the relevant title
is displayed
to confirm the user’s
choice.
Press
CLEAR to exit from
the sequence if
the wrong store
has been selected.
Otherwise,
press ENTER
to recall the stored
menu setup.
RECALL
also erases the file
contents and
sets {FILE SIZE}
to 200
blocks,
switches
the SI1287
to ’standby’, DVM
RUN to
’off, PLOT
to ’off, DVM
NULL to
’off and DVM TRIGGER
to ’single’.
{CLEAR}
> >
Remote
command is CL/. Refer
to Chapter
4,
Section 6.25.
Enter
an integer V
to erase the contents
of store ’n’.
As in
{RECALL}, the title
of
the store
(if
any)
is displayed
as a check
before any
erasure takes place.
11.8
LOCAL
>
> Remote
command
is LL. Refer
to Chapter
4,
Section
6.25.
Returns the
SI 1287
to
local
control (i.e. via
the front panel)
from the remote
state
provided Local
Lockout has not
been asserted
by an external controller.
3.32
JWS
/ SI1287 User Guide ! Issue
AF