11.5
BIAS REJECT
> >
Remote command
is RR/. Refer to Chapter
4,
Section 6.12.
Rejects the d.c. component
from the SI1287 voltage and current outputs (as
set up
via the CONDITION
menu).
[fixed] rejection
set up: key has ’on’/’off action.
[auto] rejection
set up: switching ’on’ (with the SI1287 is ’on’ mode)
causes the d.c. component of the measurement presently
being
made to be evaluated, stored, and rejected
from all subsequent
measurements, until
BIAS REJECT is switched ’off.
To re-evaluate the biases, switch ’off then
’on’ again.
11.6
DVM RUN
> >
Remote command is
RUT
Refer
to Chapter
4,
Section 6. 16.
Operates
the DVMs according to what has been selected
via the DVM
(TRIGGER) menu. Pressing the DVM
RUN key has no effect if [external] or
[sync] have been
selected for the TRIGGER menu. The LED
on
the key
glows
only whilst a measurement
is
actually taking
place; this includes measurements
triggered externally
or during a sweep cycle.
11.7
STORE/RECALL
Allows
up to
ten
menu selections
to be
stored and
recalled. Each stored selection
may be titled with
a string of up to 20 characters and spaces chosen from
the
following:
ABCDEFGHIJKI.MNOPQRSTUVWXYZabcdefghijklmnopqrstuvwxyz
%+-,7:0123456789
Note:
This facility shares memory with the history file, so its
use
affects
the
(FILE SIZE).
{STORE}
>
>
Remote command is SRI,S. Refer to Chapter
4,
Section 6.25.
Enter an integer ’n’ to
copy
existing
menu selections to store number ’n’. The
sequence after selecting
STORE is:
NUMBER RECALL
{
STORE
( ) }
(e.g.)
8
{
STORE
(8)
}
ENTER
{
STORE 8
<_ >
}
Use the
DIGIT/[
]
keys
to
move
the cursor, and the rotary knob to select the
character required
at each cursor position. Finally, press ENTER to store
the
existing menu
setup.
JWS
/ S1 1287 User Guide
/
Issue AF