S6 Portable Digital Color Doppler Ultrasound System
Service Manual
9.3.3.1 Connection through Router
You can set up the local network and DICOM functions in DICOM setting
interface.
1. Follow instructions in 9.3.2.3 to enter the DICOM setting interface
window.
Attention!
• Check/Uncheck the box next to the ezNetwork to en-
able/disable network connection.
• Network connection must be enabled in order to con-
nect the S6 system with PC or network printer.
2. Connect the router to the PC with an Ethernet cable complying with
T568B standard. Get instructions from the router manual to set the
LAN port IP address to 192.168.254.1, subnet mask to 255.255.255.0,
and turn the DHCP service on.
3. Open the computer’s network setting window. Set the PC’s IP address
to 192.168.254.X. X can be any number from 1 to 255, but not 1
or 182 (The two addresses are already taken by the router and the
S6 system.) Set subnet mask to 255.255.255.0. Set both default
gate address and first DNS sever address to the router’s IP address
(192.168.254.1).
4. Change the computer’s workgroup to "MSHOME" (Right click on My
Computer icon, click Properties-Computer Name-Change.)
5. Connect the S6 system to the router with an Ethernet cable com-
plying with T568B standard. Turn on the S6 system, wait until the
EXAM screen appears (It takes about 5 minutes). Now open the
Network Place window on the PC, the S6 system should have been
shown as a computer named "SonoXXX". If it is not shown, type
//:192.168.254.182 in the address bar, and press Enter, available con-
tents in the S6 system will be displayed in the window.
9.3.3.2 Direct Connection
Use crossover LAN cable instead of normal Ethernet cable. The crossover
cable should have pinouts as shown in Figure 9.3.3.2.
1. Follow instructions in 9.3.2.3 to enter the DICOM setting interface
window. Check the box next to "ezNetwork" to enable network con-
nection.
P/N: 4720-0034-01A
9-8