58 Chapter 6: Troubleshooting and Maintenance
Troubleshooting
Maintenance
This chapter is intended to assist in effective cleaning and disinfection. It is also intended to protect
the system and transducers against damage during cleaning or disinfection. Use the
recommendations in this chapter when cleaning or disinfecting your ultrasound system, transducer,
and accessories.
Use the cleaning recommendations in the peripheral manufacturer’s instructions when cleaning or
disinfecting your peripherals.
For more information about cleaning or disinfection solutions or ultrasound gels used with the
transducer, contact SonoSite or your local representative. For information about a specific product,
contact the product manufacturer.
There is no recommended periodic or preventive maintenance required for the iLook system,
transducer, or accessories. There are no internal adjustments or alignments that require periodic
testing or calibration. All maintenance requirements are described in the iLook User Guide and
Service Manual. Performing maintenance activities not described in the iLook User Guide or Service
Manual may void the product warranty.
Contact SonoSite Technical Support for any maintenance questions.
Touch screen is not accurate. Calibrate touch screen.
A maintenance icon
displays on the system screen
.
This icon indicates that system maintenance may be required.
Record the number in parentheses on the C: line and contact
SonoSite or your SonoSite representative.
The thermal index (TI) and
mechanical index (MI) are not
displayed on the screen.
The system and transducer combination do not exceed an MI
or TI of 1.0 in any operating mode.
Guideline is not displayed
(L25 only)
Ensure guideline is turned on.
Table 1: Troubleshooting (Continued)
Symptom Solution