Chapter 6: Troubleshooting and Maintenance 59
Troubleshooting
Recommended Disinfectant
See the Table 2, “Disinfectant Compatibility with iLook System and Transducer” on page 65.
Safety
Please observe the following warnings and cautions when using cleaners, disinfectants, and gels.
More specific warnings and cautions are included in the product literature and in the procedures
later in this chapter.
Warning: Disinfectants and cleaning methods listed are recommended by SonoSite for
compatibility with product materials, not for biological effectiveness. Refer to the
disinfectant label instructions for guidance on disinfection efficacy and appropriate
clinical uses.
The level of disinfection required for a device is dictated by the type of tissue it will
contact during use. Ensure the disinfectant type is appropriate for the type of
transducer and application. For information, see the disinfectant label instructions
and the recommendations of the Association for Professionals in Infection Control
and Epidemiology (APIC) and FDA.
To prevent contamination, the use of sterile transducer covers and sterile coupling
gel is recommended for clinical applications of an invasive nature. Do not apply the
transducer cover and gel until you are ready to perform the procedure.
Caution: Repeated, long-term exposure to coupling gel can damage transducers.
Some transducer sheaths contain natural rubber latex and talc, which can cause
allergic reactions in some individuals. Refer to 21 CFR 801.437, User labeling for
devices that contain natural rubber.