CRANEX
®
Novus e 7. Replacing parts
Service manual 97
L1800 Generator board - additional instructions
After replacing the board use the Service Assistant to
recalibrate the unit:
- calib
Refer to section 5. Service Assistant and Service
Functions.
L1700 Connector board - additional instructions
After replacing the board use the Service Assistant to
recalibrate the unit:
- calib
Refer to section 5. Service Assistant and Service
Functions.
L1200 CPU board - additional instructions
After replacing the board use the Service Assistant to
recalibrate the unit and the power supply limits and
reset the serial number.
- calib
- mainscalib
- change the IP address, if required
- update the FW, if required
- serno
Refer to section 5. Service Assistant and Service
Functions.
L1500 Power Supply board - additional instructions
After replacing the board use the Service Assistant to
recalibrate the unit.
- mainscalib
Refer to section 5. S2terminal and service com-
mands.
L1910 PFC board - additional instructions
After replacing the board use the Service Assistant to
recalibrate the unit.
- mainscalib
Refer to section 5. Service Assistant and Service
Functions.