Novus e 5. S2terminal and service commands
Service manual 79
5. S2terminal and service commands
5.1 Using the S2terminal
The S2terminal utility includes a number of commands
that allow the unit to be tested and con gured during
installation, set up and service.
To open the S2terminal:
1. Make sure that the dental imaging program being
used is closed.
2. The PC's IP address must be in same subnet. If not,
modify it according to installation manual
3. Create a new folder in the PC connected to the unit
and copy the s2terminal les into the new folder.
4. Open the Command Prompt. (Start\Programs\Ac-
cessories).
5. In the Command prompt key in cd, space and then
the path where the s2terminal les are installed
(e.g. c:\s2terminal).
_________________________________________________________________
cd c:\s2terminal
_________________________________________________________________
Press ENTER.
6. Key in s2terminal, space and then the IP address
of the unit. The IP address can be displayed by
touching the settings button on the GUI.
_________________________________________________________________
s2terminal 194.9.227.251
_________________________________________________________________
Press ENTER to open the s2terminal and make a
connection to the unit.
7. To enter the SERVICE state key in "s" and press
ENTER.
_________________________________________________________________
s
_________________________________________________________________