CRANEX
®
Novus e 3. Circuit Boards - L1800, Generator Board
Service manual 43
3.6 L1800, Generator Board
DANGER: HIGH VOLTAGE
WARNING:
Do not touch the Generator Board until the capaci-
tors have discharged. After switching the unit off
wait 10 minutes for the capacitors to discharge.
When LED H1 goes out the capacitors are dis-
charged.
L1800 - Location
In rotating unit on the tubehead side. To access remove
the tubehead external cover (see section 7.1).
L1800 - Field replaceable parts
None