66 Service manual
4. Troubleshooting CRANEX
®
Novus e
Error E8 (File system failure)
CAUSE
GUI graphics are corrupted.
SOLUTION
Update the GUI graphics.
Error E16 (Z lift not allowed due to duty cycle)
CAUSE
Z-motor over heated.
SOLUTION
Allow motor to cool down.
Error E18 (Display not connected)
Check Service Assistant for information about this
error code
CAUSE A
Cable from L1500 to L900 faulty.
SOLUTION A
Replace cable.
CAUSE B
L1500 faulty.
SOLUTION B
Replace L1500.
CAUSE C
Cable between L1200 and L900 faulty.
SOLUTION C
Replace cable.
CAUSE D
Cable between L900 and LCD display faulty.
SOLUTION D
Replace cable.
CAUSE E
L900 faulty.
SOLUTION E
Replace L900.
CAUSE F
LCD display is faulty.
SOLUTION E
Replace LCD display.