76 Service manual
4. Troubleshooting CRANEX
®
Novus e
Error E12 (CCD line failure)
CAUSE
The image data ow from the CCD sensor to L1200
was interrupted during the exposure.
SOLUTION
Refer to the trouble shooting information given in error
code E11.
Error E13 (CCD con guration checksum failure)
CAUSE
L1200 CPU failed to read CCD con guration
parameters.
SOLUTION
Reboot the unit. If the problem persists, change the
CCD sensor.