CRANEX
®
Novus e 5. S2terminal and service commands
Service manual 83
conf display summary of con guration data
Displays all the current con guration parameters of the
unit. The gures in brackets (1) are factory settings.
If you wish to restore the factory settings use the re-
store command.
NOTE:
default program 1 = adult pan, 2 = child pan, 3 = TMJ
and 5 = Test.
confccd display summary of CCD con guration data
Displays the factory and current con guration param-
eters of the CCD sensor. These values are for informa-
tion only.
dap dose area product calibration value [<value>]
Displays the dose area product (DAP) calibration value
Use values 8mA, 70kV and 1sec.
FACTORY USE ONLY!
dc takes dark current image (dc = dark current)
Is used to check if the CCD sensor is working.
Press the "Return" key to start the check.
During the check a series of values will appear in the
service assistant window.
When the check is complete the image will be trans-
ferred.
When the sensor is working correctly the vertical centre
of the image will be slightly gray because of noise.
NOTE:
If an image does not appear, then the CCD sensor is
faulty.
demo enter demo mode
Enters the unit demonstration mode.
X-rays shall be disabled and unit shall be in ready state
even without computer.
This is for exhibition use.